Contact|Site Map|Help||Japanese

2D SPR imaging equipment

image SPR area sensing
Surface Plasmon resonance (SPR) is a highly sensitive method for measuring bio-materials (protein etc) without a marker molecule (fluorescent material) while also detecting the combination reaction rate.
2D SPR imaging equipment is widely noticed as a Multi protein chip and SPR microscope.


Feature

  • Wide measurement area :
    You can measure the SPR phenomenon of a wide area at once using parallel light.
  • Easy handling :
    Sample handling is easy because the measurement chip is face up.

    image

  • Possible to measure multi-channel chip :
    Multi-channel measurement is possible using patterned chip.

    Multi-channel measurement example Y-shape flow measurement
    image image


Specifications

Instruments
Detection Method Surface Plasmon Resonance (SPR) phenomenon
SPR configuration Krechmann Configuration
Measurement range 53 ± 4°
Mesurement chip size 18 × 18 × 0.5t mm
Measurement area Depends on specification of expansion lens before CCD (standard : 5 × 5 mm)
Thermal correction none

Main Body
Light source Switch 3 types LED with optical fiber
Prism material S-TiH6
Poralization filter P/S automatically swith
Detector Cooling CCD
Size 60 × 27 × 25 cm
Weight 10kg
Power source 100V

Option
Au chip Spattered Au on glass chip
Patterning chip Photo-resist patterned on Au chip (Special order)
PDMS flow chip PDMS flow cell is attached on glass chip (Special order)
Flow cell Tool for flow measurement using Au chip

* This product is an outcome of joint research with Keio Univ. NTT and NTT-AT under KAST.